2A, as you can see, when switching to range A, the red probe needs to be plugged into port 10A which has a 10A fuse to protect the meter. When switching to either range µA or mA, the red probe needs to be plugged into the mA port which has a 600mA fuse to protect the meter, Fig 2B, 2C. Assuming the internal resistances of the current mode are: range A = 0.010; range mA = 10; range μA = 100.

(a) Is the connection in Fig. 3 correct? If not, draw the right connection.Please note the polarity. (b) How should you setup the multimeter for this current measurement? (c) If you do connect the multimeter in current mode as Fig. 3. What will happen if the multimeter is set as Fig. 2A? (d) For (c), what if it is set as Fig. 2B? (e) For (c), what if it is set as Fig. 2C?