Question

Seven blanks (of carbon free water) are measured using a total organic carbon analyzer.Measurements have a concentration range of 0 to 0.55 mg/L with a mean of 0.29 mg/L and

a standard deviation of 0.30 mg/L. What is the instrument detection limit (IDL) for this instrument at 95% confidence?